Metabolomics processing, standards and feature interpretation
Practical 17: Genomics, systems biology and multi-omics
Learning objectives
Evaluate a multiplicative-drift correction; preserve feature identity and measurement units; distinguish detected features from assigned metabolites.
Biological question and data
How does an internal standard help distinguish instrument drift from a supplied concentration pattern?
The first calculation has an ideal shared drift factor affecting analyte and standard equally. The fixture metabolite table is already log2 arbitrary abundance; its feature labels do not identify actual compounds.
No biological download is required. Numerical inputs are hypothetical, not observed animal data. The runnable examples use base R. Count and abundance units are specified where used.
1. Test an ideal measurement correction
An analyte and a standard are multiplied by the same hypothetical response factor.
The difference describes the hypothetical panel. A mass-spectrometry feature cannot be assigned to a pathway merely because its group contrast is large.
4. Inspect a calibration range
A hypothetical linear standard curve can convert a signal to concentration only under its assumptions. Show whether an unknown lies inside the supplied range.
A correct drift ratio does not identify the compound. Conversely, a confirmed identity does not make an arbitrary signal an absolute concentration. Keep calibration, internal-standard assumptions and identification confidence as separate fields.
Use blanks to investigate background and pooled QC to investigate acquisition drift. These controls address technical variation, while independent animals address biological variation. If acquisition order follows treatment, correction can be difficult to distinguish from removing the biological comparison.
Interpretation challenge
An estimate outside the highest standard looks numerically precise. Is it a validated concentration?
Suggested reasoning
No. It is an extrapolation under the fitted model. Assess the assay range, matrix effects, uncertainty and additional standards before treating it as validated quantification.
Interpretation and limits
Real workflows need blanks, technical QC, drift and batch assessment, detection/identification confidence and justified normalization. The EMBL-EBI metabolite-identification material explains why chemical assignment is separate from signal detection.
Exercises
Why did the imperfect correction fail?
Would another log transform be appropriate for M?
Can an unresolved feature support a named pathway claim?
Suggested answers
The standard did not share the same response factor as the analyte.
No; M already represents log2 abundance.
Not without justified chemical identification or carefully qualified evidence.